BA100
VECTOR SIGNAL ANALYZER

Specifications of Vector Signal Analyzer
Function
Sensitivity
Demodulator
Low, Medium and High
Digital signals, FM, NB-IoT and LTE (AM, GSM, and WCDMA will be supported in the future)
Frequency
Frequency Range
Frequenc Reference
Frequency
Readout Accuracy
Frequency Span Accuracy
Sweep Time
9kHz to 6GHz
Aging rate: ±1ppm
±( readout frequency * Frequency Reference + RBW mid-value + 0.5 * horizontal resolutionF )
±1%
1.1ms to 1600s
2.69ms to 1600s, zero span
Resolution bandwidth
RBW Range
RBW Accuracy
10Hz to 5MHz, (1-2-3-5-10 Sequence)
RBW≥1MHz,±10%
RBW<1MHz, ±2%
Amplitude
Measuring Range
Input Attenuator Range
Display Average Noise Level to +20dBm
0 to 30dB, 1dB Step
Maximum Safe Input Level
Sensitivity: +30dBm (Low)
Sensitivity: 0dBm (Medium)
Sensitivity: -20dBm (High)
Reference Level Range
Amplitude Accuracy
RBW Switching Uncertainty
Input Attenuator Uncertainty
Accuracy of Reference Level
Display Average Noise Level (DANL) @1GHz
Residual Response
-140 dBm to +20dBm
-190dBm to +70dBm (Ref level offset: ON)
±1.5dB (ATT set to 0dB, Input Signal: -5 to -30 dBm; Detector set to Positive; Sensitivity set to low; RBW Auto-Coupled, all other settings Auto-Coupled, 23±5℃, half hour warm-up required)
±0.3dB
±0.6dB
Reference level: ≥-60dBm, ±0.8dB
Input Terminated, Detector set to Positive, Trace Average set to 1000, Span set to 50kHz, Ref set to -100dBm, all other settings auto-coupled, 23±5 ℃. Normalized to 1 Hz RBW
Sensitivity: Low
-131dBm/Hz (typically -133dBm/Hz)
Sensitivity: Medium
-149dBm/Hz (typically -151dBm/Hz)
Sensitivity: High
-166dBm/Hz (typically -168dBm/Hz)
-75dBm
Input-Related Response
9kHz to 700MHz
<-70dBc
700MHz to 900MHz
<-46dBc
900MHz to 1.3GHz
<-42dBc
1.3GHz to 2.2GHz
<-46dBc
2.2GHz to 2.7GHz
<-53dBc
2.7GHz to 2.9GHz
<-42dBc
2.9GHz to 3.3G
<-38dBc
3.3GHz to 6.0GHz
<-53dBc
Second Harmonic Distortion
Third-Order Intercept (TOI)
P1dB
Phase Noise @1GHz
1.6GHz: -70dBc
+15dBm (-10dBm tones, 1MHz apart, Sensitivity set to low, Ref set to -10 dBm)
+5dBm (nominal)
-95dBc/Hz, @10kHz (typically -98dBc/Hz)
-123dBc/Hz, @1MHz (typically -125dBc/Hz)
Storage
Maximum storage depth
Data format
1Gbit
I/Q two-way, 16bit
General
Windows7, Windows 10
OS for Software
Power Supply Current
Temperature
2A Max
Operating: 0°C to 50°C
Storage: -20°C to 70°C
Connect interface
Dimension and Weight
RF output: N-type, 50 Ω
PC connect:USB type-C
Power connect: DC12V
180mm×50mm×290mm, 1.8kg
Technical specifications
This technical specifications include the influence of probability distribution, measurement uncertainty and environmental factors on the instrument performance. It guarantees the performance under the following conditions.
-
The instrument is ON and warmed up for 30 min
-
The instrument internal reference signal is applied