top of page
BN1000
Benchtop Vector Network Analyzer VNA


Specification for
Benchtop Vector Network Analyzer VNA
Measurement Range
Product Model
Impedance
Test Port Connector
Number of Test Ports
Frequency Range
Frequency accuracy
Frequency Resolution
Measurement Points
Measurement Bandwidths
Dynamic Range
Measurement Parameters
BN1000
50Ω
N-type, female
2/4
300kHz ~6.5GHz (option 265/465)
300kHz ~8.5GHz (option 285/485)
±5ppm(option 0.05ppm)
1Hz
2~20,001
1Hz to 2MHz
97 dB, typ. 115 dB (100kHz ~ 300kHz)
112 dB, typ. 115 dB (300kHz ~ 10MHz)
125 dB, typ. 130 dB (10MHz ~ 6GHz)
124 dB, typ. 129 dB (6GHz ~ 7GHz)
123 dB, typ. 128 dB (7GHz ~ 8.5GHz)
S11, S21, S12, S22
Effective System Data①
Effective Directivity
Effective source match
38~49dB
37~49dB
①Applies over the temperature range of 23°C ± 5°C after 40 minutes of warming-up, with less than 1°C deviation from the full two-port calibration temperature, at output power of 0 dBm and IF bandwidth 10 Hz
Measurement Accuracy
Measurement Accuracy
Accuracy of Transmission Measurements (magnitude / phase)
+5dB to +15dB
-50dB to +5dB
-70dB to -50dB
-90dB to -70dB
0.2dB/2º
0.1dB/1º
0.5dB/3º
2.5dB/8º
Accuracy of Reflection Measurements (magnitude / phase)
-15dB to 0dB
-25dB to -15dB
-35dB to -25dB
0.4dB/3º
1.0dB/6º
3.0dB/20º
Trace Stability
Trace Noise Magnitude (IF bandwidth 3 kHz)
2mdB rms
Temperature stability
0.01dB/℃
Measurement Speed
Measurement Time Per Point
42μs
Source to Receiver Port
Switchover Time
<10ms
Sweep points (IFBW 30kHz)
51
201
401
1601
Without calibration (300kHz~10MHz)
51points: 13ms
201points: 52ms
401points: 104ms
1601points: 413ms
Full 2-port calibration(10MHz~6.5GHz)
51points: 34ms
201points: 73ms
401points: 125ms
1601points: 434ms
Test Port Output
Match (W/O System Error Correction)
18dB
Power range (Option -70~+10dBm)
﹣50dBm to ﹢5dBm(100kHz ~ 300kHz)
﹣50dBm to ﹢10dBm(300kHz ~ 7GHz)
﹣50dBm to ﹢8dBm(7GHz ~ 8.5GHz)
Power accuracy
±1.5 dB
Power resolution
0.05dB
Test port input
Match (W/O System Calibration)
0.05dB
Max input level
﹢26dBm
Max input voltage
﹢35V
Noise level
﹣97dBm/Hz(100kHz~300kHz)
﹣107dBm/Hz (300kHz~10MHz)
﹣123dBm/Hz (10MHz~5GHz)
﹣124dBm/Hz (5GHz~6GHz)
﹣120dBm/Hz (6GHz~8.5GHz)
Other parameters
Display
External Trigger Input Connector
External Trigger Ouput Connector & External Reference Input
12.1 inch TFT color LCD, touch screen
BNC female, Input level range: 0 to +5 V
BNC female; 10 MHz; 2 dBm ± 2 dB
Video Output
USB Connector
LAN Connector
Operating Temperature Range
Storage Temperature Range
Working humidity
Working pressure
System calibration interval
Power supply
Power Consumption
Dimensions (W × H × D) mm
Weight
DP, HDMI
Female; provides connection to printer,
USB storage
10/100/1000 Base T Ethernet, 8-pin
+5°C to +40°C
-20°C to +60°C
90% (25°C)
84 to 106.7 kPa
3 years
220 ± 22 V (AC), 50 Hz
130W
440 ×230 ×360
15.5 kg
ORDERING LIST FOR VNA
Console
BN1000-265
BN1000-465
BN1000-285
BN1000-485
2-port 300kHz~6.5GHz VNA
4-port 300kHz~6.5GHz VNA
2-port 300kHz~8.5GHz VNA
4-port 300kHz~8.5GHz VNA
Calibration module
SK-CAL-NM_60
High precision, 6.5G N-type (M) calibration kit T-shaped 4 pieces set
SK-CAL-NF_60
SK-CAL-SMAM_60
High precision, 6.5G N-type (F) calibration kit T-shaped 4 pieces set
High precision, 6.5G SMA (M) calibration kit 4 pieces set
SK-CAL-SMAF_60
SK-CAL-NM_90
SK-CAL-NF_90
High precision, 6.5G SMA (F) calibration kit 4 pieces set
High precision, 9G N-type (M) calibration kit T-shaped 4 pieces set
High precision, 9G N-type (F) calibration kit T-shaped 4 pieces set
SK-CAL-SMAM_90
SK-CAL-SMAF_90
High precision, 9G SMA (M) calibration kit 4 pieces set
High precision, 9G SMA (F) calibration kit 4 pieces set
RF cable
T5_RFCAB-NmNm_60101
6.5G high precision 50Ω N-N cable
T5_RFCAB-NmSMAm_60101
T5_RFCAB-NmNm_90101
T5_RFCAB-NmSMAm_90102
6.5G high precision 50Ω N-SMA cable
9G high precision 50Ω N-N cable
9G high precision 50Ω N-SMA cable
Other options
BN1000-010
Time domain option
BN1000-1C5
Frequency rage expansion 100kHz~300kHz
BN1000-1E5
High stability clock option 0.05ppm
BN1000-1F5
Fixture circuit simulation function
BN1000-710
Power range expansion -70~+10dBm
Contact us to get more detail information about BN1000 Vector Network Analyzer